WT1800E Measure Start/Stop and EXT_Start

The Measure Start / Stop signal of the BNC connector is for connecting two WT1800E and synchronizing by master / slave operation. By setting the update rate and trigger mode to the same setting as master, slave can accurately calculate display update and integration. To start, stop, or other control integration with external signals, use the remote control signal (/EXT_START, /EXT_STOP, etc.) instead of the Measure Start / Stop signal. However, in this case, the /DA option is required. Please note that the only actions that can be executed during integration execution are stop / hold / reset, and the operation when the Start command is input again during integration execution is not regulated.