Precision Current Measurement: A Comprehensive Guide

Introduction

Designing an instrumentation system for high current measurement requires careful consideration of the trade-offs associated with each type of sensing device. The purpose of this application note is to help engineers understand the sensing choices available and the corresponding trade-offs with each technology. This guide also explores methods for high precision current measurement and high DC current measurement to ensure engineers select the most appropriate technology for their needs.

What is electrical current?

Electricity is the movement of negatively charged electrons in a conductor from a region of high electron density to a region of low electron density. The difference in electric potential between these regions is known as voltage (measured in volts). This provides the electromotive force to move electrons.

The rate of charge flow carried by these electrons per second is what is known as electric current (measure in amperes), while the opposition to this current flow is known as a resistance (measured in Ohms).

Electric current is by convention said to flow from a region of high electric potential to a region of low electric potential, opposite to the direction of electron flow.
 

High Current Measurement App Note | Components of Electrical Circuit | Yokogawa Test&Measurement
Figure 1. Basic components of electrical circuit.
 

Measuring current

As shown in Figure 1, current measurement is a series measurement of electron flow. There are two ways to make this measurement with a current sensor:

  • Direct measurement
    Accomplished by passing the current directly through a sensing device itself (shunt-type measurement).
  • Indirect measurement
    Accomplished by passing a current-carrying conductor through the aperture of a current-sensing device, such as a clamp or current transformer.

For high precision current measurement and high DC current measurement, indirect methods often offer better performance and safety.

This application note focuses on indirect technologies as they apply to the measurement of electrical power. Listed devices target the measurement of inputs and outputs of electromechanical systems such as grid-tied inverters, variable-speed motor drives, motors, chargers, generators, appliances, and transformers. The required bandwidth of such systems is typically below 1MHz. For higher bandwidths, please refer to the high-bandwidth oscilloscope probes on the Yokogawa Test&Measurement website.

Key specifications

The use of a current clamp or current transformer greatly simplifies measuring high currents (>50A) where physical constraints (e.g., conductor sizes, insertion losses, safety) make a direct measurement through the precision internal shunt of a power analyzer, DMM, or external shunt into a data acquisition instrument impractical. This convenience comes with a cost, and system designers must educate themselves on these trade-offs in order to make the best practical engineering decision.
 

Key system design specifications

Key specifications for each device must be considered when selecting a current measurement device.

  • Purposed of the measurement
  • Budget constraint
  • Physical constraint
  • Accuracy
  • Frequency response
  • Phase shift
  • Maximum current to be measured (Peak/rms)
  • Minimum current to be measured (Peak/rms)
  • Output signal (voltage/current)

Once all of the specifications are established, the sensor technology can be identified.

The most important specification to consider is the purpose of the measurement, as this often determines the rest of the specifications.

For example, if a high current measurement is needed for benchmarking power, energy consumption, or efficiency, then the accuracy specification will dictate the appropriate technology. Likewise, if a high current measurement is needed for understanding general current consumption, waveform shape, or event capture, then a different technology would likely be appropriate.

High Current Measurement | Performance vs Tech of current transformers and clamps | Yokogawa Test&Measurement
Table 1. Performance versus technology of current transformers and clamps. (*All other technologies are current output.)
 

Choosing a sensor

Indirect current measurement relies upon sensing the magnetic field generated by a current-carrying conductor. Sensing this field is accomplished through a variety of technologies such as AC current transformers, Hall-effect sensors, Rogowski coils, and fluxgate sensors. Each one of these technologies has associated trade-offs that must be considered within the system design specifications.

General selection guidelines by technology and application include:

  • Closed Aperture vs. Clamp
    A “donut-style” or “fully-closed aperture” current transformer design generally performs better than an “open clamp” or “split core” solution. This is due to the split core’s lack of symmetry and necessary discontinuity of the magnetic core and associated windings. A split core makes the closed-loop zero flux design impossible to construct, especially if multiple cores and windings are needed for accuracy, like fluxgate sensing with closedloop compensation.
  • AC Transformers
    Purely AC current transformers are typically used in line power or constant frequency applications where the voltage waveform is of a static frequency and DC components or transient phenomena are not of concern. Many current transformer manufacturers target power line frequencies specifically (50/60Hz, 400Hz), therefore the performance specifications around those frequencies are generally acceptable (error ≤ 1%). These come in clamp or donut-style designs and are ideal for high current measurement applications.
  • Rogowski Coils
    Rogowski Coils are typically used when convenience is the deciding factor, obtaining a general wave shape for an AC or pulsed signal is desired, and accuracy is not of utmost importance. Rogowski coils are very sensitive to conductor position; this error alone can contribute up to 2% of total error excluding other sources (error >1%).
  • Hall-Effect AC/DC Clamp
    Hall-effect current clamps are typically used when convenience is important, obtaining a general wave shape for any type of AC or DC signal is desired, and high accuracy is not a concern (error =>1%).
  • Fluxgate Closed-Loop
    Fluxgate-based, closed-loop zero flux current transformer designs, such as fluxgate sensors, AC transformers, and zero flux compensation, benefit from the combination of sensing technologies. Capable of making measurements on very complex AC/DC waveforms, this design often provides the best linearity, accuracy, stability, and frequency response. These devices are typically used for benchmarking power measurements where switching waveforms are present (inverters) and errors must be minimal across a wide operating bandwidth (error ≤ 1%).
     

Sensor technologies

  • AC Current Transformers
    AC current transformers rely upon Faraday’s Law, which states that an electromotive force is generated in a coil when it is placed in a time-varying electromagnetic field. The operation of a current transformer is similar to that of a voltage transformer in that the secondary output current is proportional to the turn ratio. In some cases, the current transformer can be a direct (series) measurement or an indirect measurement (through a CT opening). Because a changing magnetic field is required, they cannot be used for DC measurements. Like voltage transformers, these sensing devices are passive and do not require excitation, consisting of wire wrapped around a magnetic core.
     
    High Current Measurement | AC Current Transformer | Yokogawa Test&Measurement
    Figure 2. AC current transformer.
     
     
  • Hall-Effect Sensors
    The Hall-effect utilizes a conductive sensing material carrying a current and placed perpendicular to a magnetic field. The magnetic field produces a Lorentz force, pushing the charge carriers to one side of the material. The resulting voltage difference across the material is proportional to the flow of the current generating the magnetic field. In a current sensor, the magnetic field is generated by the primary conductor that is set 90 degrees from a Hall sensor placed inside an airgap in the iron core. The gap in the core concentrates the magnetic field around the sensor. The resulting differential voltage across the sensing material is then amplified and conditioned to an appropriate output signal (voltage or current). The Halleffect principal applies for both AC and DC signals. Halleffect current clamps usually have mV/A or mA/A output signals and may have a range switch.
     
    High Current Measurement | Hall-Effect Sensor | Yokogawa Test&Measurement
    Figure 3. Hall-effect sensor.

  • Rogowski Coils
    Rogowski Coils consist of a helical coil of wire with a center return lead wrapped around a straight primary conductor. The change in current (derivative) through the primary conductor induces a proportional voltage in the coil. This voltage is then integrated via signal conditioning electronics to produce a voltage that is proportional to the primary current. This is why Rogowski coils require a continuously changing (AC) or pulsed current in order to produce an output. The coils typically have an air core (no solid core) and therefore are flexible. Additionally, the coil wiring (center return lead) means the construction is often a flexible loop with an open end.
     
    High Current Measurement | Rogowski Coils | Yokogawa Test&Measurement
    Figure 4. Rogowski coils.
     
  • Fluxgate Sensors
    Fluxgate sensors are a saturable inductor constructed of a magnetic core and wound coil. The sensor is excited by a periodic square wave signal that sweeps the inductor across the B-H curve, in and out of saturation, resulting in a symmetric current waveform. When an external magnetic field is introduced (primary current through an aperture of the CT), the flux density is changed and the current waveform becomes asymmetric due to the change in time to reach saturation. This asymmetry is measured via a change in duty cycle or via harmonic analysis, producing a linear output with respect to the primary current. Fluxgate sensors are capable of sensing both AC and DC currents and typically offer greater stability, resolution, and accuracy when compared to Hall sensors. Similar to a Hall clamp, a fluxgate sensor can be placed inside an airgap or constructed out of a second magnetic core placed inside of the current transformer itself.
     
    High Current Measurement | Fluxgate Sensor | Yokogawa Test&Measurement
    Figure 5. Fluxgate sensor.



    High Current Measurement | Fluxgate Sensor Without Primary Current | Yokogawa Test&Measurement
    Figure 6. Fluxgate sensor without primary current.


    High Current Measurement | Fluxgate Sensor With Primary Current | Yokogawa Test&Measurement
    Figure 7. Fluxgate sensor with primary current.
     
     

Errors

All of these technologies can have measurement errors attributed to factors such as linearity, offset, temperature, or noise.

  • Errors in AC current transformers are can be a function of manufacturing quality or residual flux.
  • Hall-effect sensors are vulnerable to errors arising from the necessity of sensing and amplifying very small voltages and the properties of the Hall sensing material (offset, temperature drift, noise).
  • Rogowski coils are prone to positioning, droop/offset, and phase delay errors.
  • Linearity error is a concern for any type of sensor, and is a function of the technology, construction, and temperature characteristics.
  • Fluxgate sensors provide more robust measurement due to a time-based measurement (duty or spectral analysis), low magnetic offsets due to the cyclic excitation, and a low temperature drift.
     

Zero flux technique

To compensate for linearity errors, operate the sensing technology in a “zero flux” condition, where the magnetic field being measured by the sensor is essentially zero. This is accomplished through a compensation winding inside of the current transformer that generates an equal-yet-opposing magnetic field to that of the primary field. This winding is driven in a closed-loop circuit formed by the sensor (fluxgate or Hall) and associated amplification circuitry as shown in Figure 8. This allows the sensor to essentially operate around a zero sensing condition (a single point), minimizing any gain errors. Offset errors can further be eliminated by applying a zero offset or a “nulling function” in the sensing instrument that the current transformer is connected to, such as a data acquisition, oscilloscope, or power analyzer.
 

High Current Measurement | Closed-Loop Hall-Effect Zero Flux CT | Yokogawa Test&Measurement

Figure 8. Closed-loop Hall-effect zero flux CT.
 
 
High Current Measurement App Note | Closed Loop Fluxgate Zero Flux CT | Yokogawa Test&Measurement
Figure 9. Closed-loop fluxgate zero flux CT.
 

Frequency response

Current sensing technology and associated electronics have limited bandwidth. However, an advantage of the closedloop zero flux configuration is that at higher frequencies, the compensation winding acts as an AC current transformer. This significantly extends the bandwidth and reduces the response time of the transducer. Essentially, the zero flux (closed-loop) current transformer design incorporates multiple current sensing techniques (e.g., AC current transformer, Hall-effect, fluxgate). As a result, zero flux designs are capable of measuring AC, DC, and complex waveforms of any shape.

High Current Measurement App Note | Frequency Response of Closed Loop Design | Yokogawa Test&Measurement
Figure 10. Frequency response of closed-loop design.
 

High Current Measurement App Note | Multi-Core Closed-Loop Fluxgate Design | Yokogawa Test&Measurement
Figure 11. Multi-core closed-loop fluxgate design.
 

Output

Yet another advantage of the closed-loop design is that the output signal is current-based. This provides a more robust signal in high noise environments. It is also the preferred signal for power analyzers, which can directly measure current with high accuracy. When dealing with any current transformer, the output is considered a constant current source. This means as current flows through the primary conductor, the secondary must never be left as an open circuit. An open circuit essentially produces infinite resistance, by Ohm’s law V = I*R, and results in a very high voltage that damages the current transformer and presents a significant safety hazard. Design systems with caution so that the current transformer does not disconnect when performing a measurement.
 

High current power measurement configuration examples

High Current Measurement App Note | Power Analyzer with Built In Power Supply | Yokogawa Test&Measurement
High Current Measurement App Note | Power Analyzer with Built In Power Supply | Yokogawa Test&Measurement
Figure 12. Power analyzer with built-in power supply.
*WT1800 - Total current consumption for IN 2000-s cannot exceed 6Arms (+2Arms secondary).
*PX8000 - Total current consumption for IN 2000-s cannot exceed 4Arms (+2Arms secondary).
 
 
High Current Measurement App Note | Power Analyzer with External Power Supply Kit | Yokogawa Test&Measurement
High Current Measurement App Note | Power Analyzer with External Power Supply Kit | Yokogawa Test&Measurement
Figure 13. Power analyzer with external power supply kit.
*Total current consumption for IN 2000-s cannot exceed 6Arms (+2Arms secondary).
*Yokogawa M114** kits include IT or IN CTs and IST power supply (banana/fork terminals not included).
 
 
High Current Measurement App Note | Power Analyzer with Built In Power Supply Integrated in the Element | Yokogawa Test&Measurement
High Current Measurement App Note | Power Analyzer with Built In Power Supply Integrated in the Element | Yokogawa Test&Measurement
Figure 14. Power analyzer with built-in power supply integrated in the element.
 

Integration considerations

After selecting the appropriate current sensing product, the integration of the device must be carefully engineered, thoroughly evaluating all output types, accuracy, measurement range, and interconnections.
 

Example use cases

Case One

A manufacturer of UPS systems is engineering an instrumentation system to monitor and make measurements on six phases of AC voltage, current, power, and one DC phase (480Vrms, 75Arms). Power and harmonic monitoring is important. However, capturing distorted wave shapes during changeover events is critical. The instrumentation system is data acquisition-based and requires the ability to easily move to different installations.

Data acquisition systems generally consist of voltage signals and are typically not highly accurate when making power measurements, so it makes little sense to select a highly accurate current device. In this case, the engineer has selected a Hall-effect-based clamp-style sensor, which is valid for any type of wave shape (AC/DC).

Engineering Considerations

  • The output voltages from Hall-effect clamps are often in mV/A, where the maximum output voltage is less than 1V. The data acquisition system should have a comparable fullscale input range. For example, utilizing a 10V input range would result in poor resolution measurement. Take into consideration the proper scaling for the instrument when using clamps that have multiple range switches (400A/40A), as changing range on the clamp requires a range change on the instrument.
  • Hall-effect clamps require power, usually supplied by a small 9V or AA batteries, USB power, or a wall-style AC-to-DC plug.
  • Many Hall-effect clamps have a BNC-style interconnection. The data acquisition system should have the same or an adapter, if banana jacks are used.
  • Hall-effect is prone to offset drift. Because of this, the nulling (zero offset) function of the probe itself and/or the data acquisition should be used before making critical measurements.
High Current Measurement App Note | Example Configuration of a Six-Phase High-Power Monitoring System | Yokogawa Test&Measurement
Figure 15. Example configuration of a six-phase high-power monitoring system.
 
 
High Current Measurement App Note | Configuration chart for Case One using DL950 ScopeCorder with Real-Time Power | Yokogawa Test&Measurement
Table 2. Configuration chart for Case One using DL950 ScopeCorder.

Case Two

A manufacturer of inverter-based motor drives is engineering an instrumentation system for benchmarking power measurements on three phases of PWM-based AC voltage, current, power, and one DC input phase (800Vrms, 1100Arms). Waveforms are of interest. However, capturing the most accurate power measurements is most critical, as the efficiencies of new inverter designs are over 90%. The instrumentation system is power analyzer-based and must provide highly accurate power, energy, and harmonic measurements.

In this instance, the engineer has selected a fluxgate-based zero flux current transformer. The fluxgate is valid for any type of wave shape (AC/DC) and provides the highest accuracy solution for PWM application requiring high bandwidths. Power analyzers are highly accurate devices, and the most precise current measurement is required to keep errors to a minimum. A low-accuracy clamp or fixed-frequency current measurement device would not be a good selection.

Engineering Considerations

  • Because this is a “donut style” design, the primary cable must be able to pass through the center aperture (check diameter). In some cases, a bus bar or smaller cable with smaller diameter insulation is necessary.
  • This is a powered device and requires a power supply with the appropriate voltage and current capacities, along with a location for mounting (rack).
  • Combining the fluxgate AC current transformer technologies in a closed-loop method, there is always a constant current output on the secondary when the primary has current flow. The current transformer cannot be disconnected when a measurement is being performed. Appropriate labels or design considerations should be made for safety and to avoid damage to the device.
  • The closed-loop sensor provides a current-based output that is less susceptible to noise. This is the preferred signal type for a high-performance power analyzer (such as the Yokogawa WT series), that employs a unique shunt design that compensates for thermal drift and ensures the most precise measurement.
  • Size the full-scale output of the secondary current appropriately in relation to the power analyzer. For example, a 2000:1 ratio for a 280Arms transformer results in a 140mA full-scale output. The power analyzer should have comparable low current measurement ranges.
  • A power analyzer with a 40A input and a 500mA minimum full-scale range is too large. A better choice is a power analyzer with a 5A range and a 5mA minimum full-scale range.
  • While closed-loop sensor designs are less prone to offset drift, the nulling (zero offset) function of the power analyzer can be used to further reduce offset error.
     
    High Current Measurement App Note | WT5000 Power Analyzer Example Configuration | Yokogawa Test&Measurement
Figure 16. WT5000 Precision Power Analyzer example configuration.
 
 
High Current Measurement App Note | WT5000 Precision Power Analyzer Configuration Chart | Yokogawa Test&Measurement
Table 3. Configuration chart for Case Two using a WT5000 Precision Power Analyzer.
 
 
High Current Measurement App Note | WT1800E Power Analyzer Example Configuration | Yokogawa Test&Measurement
Figure 17. WT1800E Precision Power Analyzer example configuration.
 
 
High Current Measurement App Note | WT1800E Precision Power Analyzer Configuration Chart | Yokogawa Test&Measurement
Table 4. Configuration chart for Case Two using a WT1800E Precision Power Analyzer.
 

Cable Length Considerations

When it is necessary to lengthen the cables of the IT, IN, or CT series current transformers, the voltage drop in power supply lines and total burden resistance seen by the transformer needs to be considered.

  • The voltage drop across the power supply cables must be <%5 of the +15V lines.
  • The total resistance seen by the secondary of the current transformer must be less than the transformers RM (burden resistance) specification. This is curve of primary current vs. resistance, with multiple operating temperatures. It is important to consider the resistance of the entire signal length from transformer to power analyzer, as well as the internal resistance of the power analyzer.

Example:

  • IN 2000-s current transformer
  • WT1800E with 5A element
  • Maximum operating current 3000Apk
High Current Measurement App Note | Lengthening IN IT Transformer Cables | Yokogawa Test&Measurement
Figure 18. Lengthening IN/IT transformer cables.
 
 
High Current Measurement App Note | WT5000 Precision Power Analyzer and IN 2000 S Specifications | Yokogawa Test&Measurement

High Current Measurement App Note | WT5000 Precision Power Analyzer and IN 2000 S Specifications | Yokogawa Test&Measurement
Figure 19. WT5000 Precision Power Analyzer and IN 2000-s specifications.

 

The specifications for the 5A module state the internal resistance is 100 mΩhm. The IN 2000-S specifications show the maximum burden resistance at 3000Apk is approximately 1 Ω @ 25C, with a transformation ratio of 2000:1 (NS), and an overhead current consumption of about 200mA (IC).

  • Size power supply cables to maintain <0.75V drop when operating at 3000Apk primary current or 1Apk of secondary current ~1.2Apk total supply current.
  • At 2000Apk primary current (1Apk secondary) the total resistance of the signal cables from the current transformer to the power analyzer must be <0.89 Ω @ 25 (1Ω-0.11Ω).

Recommendations on wire types and shielding for the power supply cables are detailed in the IST power supply user manual. For long runs, twist the signal wires from the power supply to the power analyzer to reduce the influence of noise, though bear in mind this increases the total length of the wires.
 

Combining uncertainty for power measurements

Power analyzer uncertainty is a percentage of reading plus a percentage of range error. Current transformers are a percentage of reading plus offset. The following shows methods for estimating the total uncertainty in a power analyzer and current transformer system.

Example (sum of errors/worst case):

High Current Measurement App Note | Power Measurement Uncertainty Chart | Yokogawa Test&Measurement

High Current Measurement App Note | Power Measurement Uncertainty Equations & Calculations | Yokogawa Test&Measurement
Figure 20. Estimation methods for total uncertainty in a power analyzer and current transformer system.

Related Industries

Related Products & Solutions

AC/DC Split Core Current Sensor CT1000S

Measure high currents without disassembling existing cabling. Compatible with power analyzers and waveform measurement instruments.

Digital Power Meter WT300

The WT300 offers high performance and high reliability in a compact package with excellent price value, making it ideal for multiple applications from R&D to manufacturing test. It is the fifth generation of Yokogawa's value series power meter, and the world's best selling power meter.

DL350 Portable ScopeCorder

A compact, battery-powered, touch screen instrument you can take into the field featuring the deep measurement and recording capabilities you need in the lab.

DL850/DL850V ScopeCorder

Combining the capabilities of an Oscilloscope and the characteristics of a Data Acquisition Recorder, the Yokogawa Scopecorder is a versatile tool for engineers making measurements in the industrial, transportation and energy sectors.

DL850E/DL850EV ScopeCorder

An all-in-one electromechanical measurement system offering the deepest measurement and analysis features to turn data into insights.

DL950 ScopeCorder

The Yokogawa DL950 ScopeCorder captures and analyzes a wide variety of electrical, physical sensor signals, and serial buses. It offers a unique combination of high sampling rates, for a detailed view and long recording times to monitor trends over time.

PowerViewerPlus for PX8000

The PowerViewerPlus (part #760811) enables PC-based remote control, acquisition and analysis of high-frequency and transient power signals from a PX8000 Precision Power Scope.

Precision Power Analyzer WT3000

With 0.02% accuracy and 1MHz bandwidth, the WT3000 delivers where the highest precision measurements are required. It is the industry standard for R&D work on inverters, motor drives, lighting systems and electronic ballasts, UPS systems, aircraft power, transformer testing, and other power conversion devices.

PX8000 - Precision Power Scope

  • Transient/Start-up Waveform Capture & Power Analysis
  • Harmonics, Cycle by cycle and FFT analysis

Wirepuller Controlling DL Series & WT1600

Wirepuller enables the Digital Oscilloscopes and WT1600 Digital Power Meter to be controlled from your PC via the Ethernet, USB, or GP-IB interface.

WT1600 Digital Power Meter

The High-end WT1600 is designed for a wide range of applications, from energy-saving to large load applications. The WT1600 has a six element input capability to provide maximum flexibility, and Basic Power Accuracy of 0.1% with 1 MHz Bandwidth.

WT1800 High Performance Power Analyzer

The WT1800 Power Analyzer offers maximum flexibility with up to six wattmeter elements, high bandwidths, simultaneous high speed digitizing, and wide voltage and current ranges. Although this model is still available, the newer WT1800E offers higher accuracy.

WT1800E - High-Performance

  • Power, Harmonics (THD) Measurements (up to 500th order)
  • Electrical & Mechanical Efficiency of Inverters and Motors in EV and HEV

 

 

WT1800R High-Performance Power Analyzer

  • Power, harmonics (THD) measurements (up to 500th order)
  • Electrical and mechanical efficiency of inverters and motors in EV and HEV

WT2010/WT2030 Digital Power Meters

The WT2000 digital powermeter series has been designed with emphasis on basic performance (bandwidth, accuracy, response speed, and noise immunity) from the viewpoint of measurement of electrical quantities. The broad range of functions of these power analyzers enable them to be used in various fields of applications.

WT210/WT230 Digital Power Meters

Superior cost performance and compact size make the entry level WT200 Series Digital Power Meters ideal for manufacturing and basic power measurement applications. Offered with One to Three Elements, with Basic Power Accuracy of 0.1% and 100kHz Bandwidth.

WT3000E - High Accuracy

  • High Accuracy Electrical & Mechanical Efficiency of Inverters and Motors in EV and HEV
  • IEC Harmonics (IEC61000-3-2 & IEC61000-3-12); Voltage Fluctuations and Flicker (IEC61000-3-3 & IEC61000-3-11)

 

WT300E - Economy

  • Standby Power Measurements, Energy Star®, SPEC Power® and IEC62301/EN50564
  • Evaluation and Testing of Batteries, Home Appliances and Uninterruptible Power Supplies (UPS)

 

WT500 - Mid Range

  • Power and Harmonic (THD) Measurements with Independent Range Controls
  • Efficiency measurements of AC/DC or DC/AC Inverters & Motors

 

WT5000 - Precision Power Analyzer

  • Yokogawa WT5000 precision power analyzer with superior accuracy
  • 7 swappable input elements
  • 4 motor inputs
  • Harmonics test compliant to IEC 61000-3-2

WTViewer Application Software

The WTViewer is an application software tool that reads numeric, waveform, and harmonic data measured with the WT3000E /WT3000 /WT1800E /WT1800 /WT500 Digital Power Analyzer.

WTViewer for WT210/WT230 (FREE)

Easily Acquire and Manage Power Measurement Data from Your PC

WTViewerE Application Software

WTViewerE software enables PC connectivity for Yokogawa power analyzers such as the WT5000, WT3000E/ WT3000, WT1800E/ WT1800 , WT500 and WT300E/WT300 through Ethernet, USB, GPIB or RS232. This connectivity allows users to easily control, monitor, collect, analyze, and save measurements remotely.

WTViewerEfree Software

The WTViewerEfree is free software for power analyzer models that include WT series*. The software simplifies instrument configuration and gathering of measurements.

*WT5000/ WT3000E/ WT3000/ WT1800E/ WT1800/ WT500/ WT300E/ WT300

WTViewerFreePlus for WT300/WT300E Series

The WTViewerFreePlus software captures measured numeric values, harmonic values, and waveform data. Users can view and save data on a PC using USB, GPIB, RS-232, or Ethernet.

Calibrators

  • Calibrators for full spectrum of commonly measured signals in plant operations 
  • Trusted reference for verifying industrial instrumentation

Power Analyzer Accessories

Accessories for digital power analyzers include various voltage and current transformers, clamp-on current probes, and a selection of test leads.

Power Analyzers and Power Meters